List of program instructions – Metrohm IC Net 2.3 User Manual

Page 265

Advertising
background image

6.28 844 UV/VIS Compact IC

IC Net 2.3

255

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

subpage:

Remote

0, 1, *, p

Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter

1

,

0

,

p

or

*

. For entry of the

other values, move the cursor in front
of the value to be changed and enter

1

,

0

,

p

or

*

.

Valve Inject,

Fill

Switch injection valve to "INJECT"
or "FILL" position.

ICPump on,

off

Switch on or off the high-pressure
pump
.

Flow

0.2 ... 2.5
mL/min

Set flow rate of the high-pressure
pump to the desired value.

Pmax

0.0 ... 35.0 MPa

Set maximum pressure limit for the
high-pressure pump to the desired
value.

Pmin

0.0 ... 35.0 MPa

Set minimum pressure limit for the
high-pressure pump to the desired
value.

Peristaltic on,

off

Switch on or off the peristaltic
pump
.

Wavelength_1..3

190 ... 900 nm

Set wavelength of the corresponding
measuring channel on the desired
value.

Bunching_1..3

1 ... 101 nm

Set bunching of the corresponding
measuring channel on the desired
value.

UV_Lamp on,

off

Switch on or off the UV lamp.

VIS_Lamp on,

off

Switch on or off the VIS lamp.

Zero

Record a new zero spectrum.