6.13 Metrohm 818 Solvent Delivery Unit

IC Net 2.3

141

Stop after

Stop time for all pump drives.
Entry range:

0 ... 99999 min

<Set>

Send current parameters immediately to the
pump control system. Parameters are not
stored in the system file (

*.smt

) as long as

the file is not saved.

<Start>

Start all pump drives.

<Stop>

Stop all pump drives.

Program

The

Program

tab of the

Pump Control Unit

window contains a time

program for the pump control system to define eluent gradients.

Flow

Overall flow of pump control system in
mL/min. This value is also available for the

WATCH WINDOW

. This field is only available

if the

Show flow column

option is disabled.

Entry range:

0.01 ... 20.00 mL/min

(depending

on number of pumps)

Stop pumps after the run

Stop pumps after the determination.

Program table

Number

Row number in the program table (read-
only).

Time

Time at which the flow and/or gradient is
changed.
Entry range:

0.0 ... 99999 min