6 Devices

IC Net 2.3

116

Entry field

Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection:

fill, inject, off

* only available for

733 IC Separation Center

connected to 732

Time program

The

Program

tab is only available for a

733.0XX0 Separation Center

connected to a 762 IC Interface. The

733 IC Separation Center

that is connected to a 732 IC Detector is controlled by the time
program of the detector.
On the

Program

tab of the

733.0XX0 Separation Center

window a

user-defined time program can be entered. This program is
started automatically as defined in the

Start mode

window (see

section 4.4.3) either at the moment the determination is started
(

Start with determination

) or at the moment the sample is injected

(

Start with inject

).

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see be-
low).

ENABLED

Enable program start (a disabled program
is not started).