Metrohm IC Net 2.3 User Manual

Page 218

Advertising
background image

6 Devices

IC Net 2.3

208

Entry range:

0…999 µL

, default:

30

(with air

segment)

The minimum flush volume depends on
the volume of needle and needle tubing,
flush volumes of less than twice the vol-
ume of needle and tubing will result in
decreased performance.

Inj.

volume, µL

Programmable only for

Partial

and

Pick-up

modes, is equal to the

Loop volume

in

Flushed

mode. Is loaded from the sample

queue. The maximum injection volume de-
pends on the injection mode:

Partial:

50% of the loop volume

Pick-up:

(loop volume – 3

×

needle vol

ume)/2
Entry range:

1…1000 µL

Analysis time

Time between switching the injection valve
to inject and the start of processing the next
sample. Is loaded from the sample queue.
Entry range:

0:00:00…9:59:59 h:mm:ss

Delay time

Time

between

start determination

and start

of the

Triathlon methods/User program

, useful

e.g. for baseline equilibrium after a sup-
pressor step.
Entry range:

0:00:00…1:00:00 h:mm:ss

<Set>

Apply parameter values selected under

Method values

to the

Injection method

.