Controls – Metrohm IC Net 2.3 User Manual

Page 260

Advertising
background image

6 Devices

IC Net 2.3

250

Controls

On the subtab

Controls

injection valve, peristaltic pump, high-

pressure pump, column heating and remote lines can be manu-
ally controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:

Valve

Injection

valve

<Inject>

Switch injection valve to "Inject" position.

<Fill>

Switch injection valve to "Fill" position.

Peristaltic pump

Peristaltic

pump

<On>

Start peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).

<Off>

Stop peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).

IC pump

High pressure pump

<On>

Start high pressure pump.

<Off>

Stop high pressure pump.

Flow, mL/min

Set value for flow rate of IC pump.
Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Value for maximum switch-off pressure for

high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under

IC NET / Global

preferences

).

Entry range:

0.0

...

35.0 MPa

If the

Flow

is higher than 1.5 mL/min, the

defined

Pressure max

must not exceed