6.28 844 UV/VIS Compact IC

IC Net 2.3

261

Detector settings

On the tab

Detector settings

measuring settings are defined.

Detector settings

Settings for UV/VIS detector.

VIS lamp

Display of the voltage of the VIS lamp.

T int, ms

Integration

time.

Entry range:

10

...

254 ms

T out, s

Measuring time per data point.
Entry range:

0.05

...

2.00 s

Data rate

Effective measuring time per data point.
The

Data rate

is the particular whole number

multiple of the value

T int

that is closest to

the value

T out

.

The

measurement

of a data point is explained by using an

example

:

Set values

:

T int

= 15 ms,

T out

= 0.1 s

The

calculated

Data rate

is 105 ms (the particular multiple of

T

int

that is closest to the value

T out

; in our example 7 times 15

ms). This means that the integration will be carried out 7 times

(each time for 15 ms) and then averaged to give a data point.

<Optimize>

Optimizes

T int

and adjusts the VIS lamp.

It should be done

after exchanging a lamp (UV or VIS)

if parts of the intensity spectrum are in the

overload ( > 60'000 , peaks cut )

if the maximum of the intensity spectrum of

dist. water is too low (below 33'000)

Before clicking

<Optimize>

, the flow-trough

cell should be rinsed and filled with dist.
water. There must be no air bubbles in the
flow-trough cell when clicking

<Optimize>

.