Thermostat – Metrohm IC Net 2.3 User Manual

Page 133

Advertising
background image

6.9 820 IC Separation Center

IC Net 2.3

123

Thermostat


The

Thermostat

tab of the

820 IC Separation Center

or

820.02X0 IC

Separation Center

window contains the temperature control of the

column heater.

Only available for a 2.820.02X0 Separation Center that is con-
nected either to the 819 IC Detector and installed as 820.02X0
to IC Detector
, or to the 830 IC Interface and installed as
820.02X0 to Interface.

Actual temperature

Display

of

actual temperature within the

column heater in °C.

Set point

Operating temperature of the thermostat
within the column heater in °C.
Entry Range:

10.0 … 80.0

.

<On>

Only active if the thermostat is off. Switch
the thermostat of the column heater on and
apply the operating temperature specified
under

Set point

, the button turns to

<Set>

.

<Set>

Only active if the thermostat is on and a
new operating temperature is entered at

Set

point

. Sets the new operating temperature.

<Off>

Only active if the thermostat is on. Switch
the thermostat of the column heater off.

Advertising