6.10 709 IC Pump

IC Net 2.3

127

value during operation, it is switched off
immediately. At the same time, the

Pressure

field color in the 709 settings window and in
the

WATCH WINDOW

is changed to the

value set for

Out of range

.

The

set

minimum limit value should lie far

enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is
automatically switched off. At the same
time, the

Pressure

field color in the 709 set-

tings window and in the

WATCH WINDOW

is

changed to the value set for

Out of range

.

<Set>

Send current parameters immediately to the
709 IC Pump (only available if pump is run-
ning). Parameters are not stored in the sys-
tem file (

*.smt

) as long as the file is not

saved.

<Start>

Set current parameters and start 709 IC
Pump drive.

<Stop>

Stop 709 IC Pump drive.

Links

The

Links

tab of the

709 IC Pump

is used for COM port selection

and settings (details see section 5.2.4 Links).