Outputs – Metrohm IC Net 2.3 User Manual

Page 257

Advertising
background image

6.27 861 Advanced Compact IC

IC Net 2.3

247

For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.

<Break times>

Possibility for changing the break times for
injection valve

Valve

and suppressor mod-

ule

Suppressor

.

Outputs

The

Outputs

tab of the hardware settings window defines remote

output signals to be set automatically if specific events occur.

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "IN-
JECT" position.

Fill

Switching of the injection valve to the "FILL"
position.

Suppressor step

Switching of the suppressor module to the
next position.

Pump stop (P max/min)

Pump stopped because pressure limits are
exceeded.

Hardware error

Hardware error detected at the 861 Ad-
vanced Compact IC (high-pressure pump,
injection valve, or suppressor not working
correctly).

Alarm leak detector

Leak detector has detected solvent in the

instruments interior.

Remote line

Set remote output lines 1

8.

Selection:

0

(line off, inactive, open)