KEW6310 6.4.1 Instantaneous value measurement -Ranges

KEW6310

6.18

6.4 Ranges and Over-range indications
6.4.1 Ranges

Ranges and decimal points for the measuring items will be automatically adjusted depending on the
settings for Voltage, Current Ranges and VT / CT ratio.

Voltage Range : V, Max digit : 4-digit

(V Range) x (VT ratio) x (120%)

Decimal point & Unit

1.8 ~ 9.999 V

9.999 V

10 ~ 99.99 V

99.99 V

100 ~ 999.9 V

999.9 V

1 ~ 9.999kV 9.999kV

10 ~ 99.99kV 99.99kV

100 ~ 9.999kV 999.9kV

1 ~ 9.999 MV

9.999 MV

10 ~ 12.0 MV

12.00 MV

Current Range : A, Max digit : 4-digit

(A Range) x (CT ratio) x (120%)

Decimal point & Unit

1.2 ~ 9.999 mA

9.999 mA

10 ~ 99.99 mA

99.99 mA

100 ~ 999.9 mA

999.9 mA

1 ~ 9.999 A

9.999 A

10 ~ 99.99 A

99.99 A

100 ~ 999.9 A

999.9 A

1 ~ 9.999kA

9.999kA

10 ~ 99.99kA

99.99kA

100 ~ 999.9kA

999.9kA

1 ~ 9.999 MA

9.999 MA

10 ~ 36.00 MA

36.00 MA