KEW6310 4.2.2 Measurement Setting (QUALITY-Transient Measurement)
KEW6310
4.50
Setting for hysteresis
1 ~ 10% (can be set by 1%)
* Default value (or after system reset) : 5%
* Setting procedure is same to that for Hysteresis Setting for Swell, Dip, Int measurement.
Refer to the procedure described in the preceding pages.
Setting for trigger point
Past: 1 ~ 200 (can be set by 1)
Next : 200 ~ 0 (can be set by 1)
* Default value (or after system reset) : 100
* Trigger to start and stop recording when a preset threshold exceeded will be decided based on
the number of recorded data.
* Setting procedure is same to that for Trigger Point Setting for Swell, Dip, Int measurement.