5 capacitance calculation, 8 other specifications, Kew6310 – KYORITSU 6310 Instruction Manual User Manual

Page 291: Cos θ, 1) ac power supply, 2) dc power supply, 3) battery charge

Advertising
background image

KEW6310 16.7.5 Capacitance calculation

KEW6310

16.8

16.7.5 Capacitance calculation

Display item

Same to W Range (except for the change from PA to C)

Save item

(Measurement data at W Range) + (calculated capacitance value)

Formula

[

]

[ ]

F

V

f

P

k

P

C

μ

θ

θ

θ

θ

)

1

cos

1

1

cos

1

(

2

10

var

)

1

cos

1

1

cos

1

(

0

2

1

2

2

9

0

2

1

2



Ч

Ч

Ч

=



Ч

=

π

C

: Capacitance needs for improvement

P

: Load power [kW]

:

Frequency

V :

Voltage

1

cos

θ

: Measured power factor

0

cos

θ

: New power factor (target)

16.8 Other specifications

(1) AC power supply

Voltage range

AC100 ~ 240V±10%

Frequency 45

~

65Hz

Power consumption

20VA max

(2) DC power supply

Dry

battery

Rechargeable

battery

Type Alkaline

(LR6)

Ni-MH(HR-15-51)

Rated voltage

DC9V (=1.5Vx6)

DC7.2V (=1.2Vx6)

Current consumption

500mA typ.(@9V)

560mA typ.(@7.2V)

Possible measurement time

Backlight ON: 1 hour
Backlight OFF: 2 hours
(ref. at 23ºC)

Backlight ON: 2 hours
Backlight OFF: 5 hours
(ref. at 23ºC after full-charge)

(3) Battery charge

Charging voltage

approx 9V

Charging current

approx 400mA

Charging pattern

Charging pattern is as follows to control whole current consumption.

Pattern Charging

Pause

Total

charging time

I. Power ON,LCD_ON 0.7 4.3

48

II. Power ON,LCD_OFF

2.1 2.9 14

III. Power OFF 4.2

0.8

7

[min] [hour]

Start charging

Following should be completely met.

- Supply of power from AC power supply
- Selector switch is set to “Rechargeable battery” position.
- Operation to start battery charge

Finish charging

Battery charge stops if any of following is met.
<For Pattern I, II>

(1) power from AC power supply is stopped,
(2) selector switch is set to “Dry battery” position,
(3) 48hours later from the start of battery charge,
(4) Battery voltage becomes lower than that checked at previous pause period,
(5) charging voltage is 9.5V or more (batteries are removed),
(6) specific charging cycle is exceeded.

<For Pattern III>
Battery charge stops if Any of (1), (4), (5), (6) is met.