Setting for int (short interruption) – KYORITSU 6310 Instruction Manual User Manual

Page 83

Advertising
background image

4.2.2 Measurement Setting (QUALITY-Swell, Dip, INT Measurement) KEW6310

KEW6310

4.42

Setting for int (short interruption)

5 ~ 98% (can be set by 1%)

* Default value (or after system reset) : 10%

1

Press the

Cursor

Keys and select [Short interruption], and then press the

ENTER

Key.

2

Press the

Cursor

Keys and alter values, and then press the

ENTER

Key.

Lower limit varies depending on the selected reference voltage. Alter the reference voltages to
change the lower limit.

Box with

▲▼

mark appears

at the rightmost digit.

Selected int value is

displayed.