Auto setting – KYORITSU 6310 Instruction Manual User Manual

Page 52

Advertising
background image

4.2.1 Basic Setting KEW6310

4.11

KEW6310

<< Auto Setting >>

Model name of the Clamp sensor connected to the Current Terminal of the instrument is detected auto-
matically at Auto setting mode. Setting for [Wiring] should be done to advance Auto setting.

1

Confirm that settings for [Wiring] are made, and then press the

F2

Key.

The max measurable values on Clamp sensor are reflected in setting for [A Range].
[CT ratio] is automatically set to 1.00.
For [Filter], bars are displayed when the detected sensors are MODEL812X series and OFF is
displayed when the sensors are MODEL814X series.

Setting will be changed if new sensors are detected at powering on the instrument.

Pressing the

F2

Key initiates

auto setting for Clamp sensor.

Connected Clamp sensors are automatically

detected, and settings for [A Range], [CT ratio]
and [Filter] are made automatically.