KEW6310 14.1 Input/ Output terminals

KEW6310

14.2

[Input terminal]


Capable of measuring and recording DC voltage signals.
Number of Ch: 2ch
Input resistance : approx 225kΩ

CAUTION


Roots of the L terminals for each Ch are integrated. Never connect inputs with various grand levels to
the terminal at the same time.


[Output terminal]


Capable of generating outputs when events occur during measurements below.

Measurement menu

Conditions for Output : Lo

Remarks

Demand

(Predicted value) > (Target value)

------

Harmonic

exceeding preset allowable
range

Lo output; when an alowable
range is exceeded at any Ch.

Swell/ Dip/ Int/ Transient/
Inrush

new event is added and
displayed on the LCD

Lo is kept for 1 sec, Hi is
restored

Unbalance rate

exceeding preset threshold

------



Output format : Open collector output
Max input : 30V, 50mA, 200mW
Output voltage : Hi – 4 ~ 5V

Lo – 0 ~ 1V



+5V

H

L

OUTPUT