Input terminal, Caution, Output terminal – KYORITSU 6310 Instruction Manual User Manual
Page 277

KEW6310 14.1 Input/ Output terminals
KEW6310
14.2
[Input terminal]
Capable of measuring and recording DC voltage signals.
Number of Ch: 2ch
Input resistance : approx 225kΩ
CAUTION
Roots of the L terminals for each Ch are integrated. Never connect inputs with various grand levels to
the terminal at the same time.
[Output terminal]
Capable of generating outputs when events occur during measurements below.
Measurement menu
Conditions for Output : Lo
Remarks
Demand
(Predicted value) > (Target value)
------
Harmonic
exceeding preset allowable
range
Lo output; when an alowable
range is exceeded at any Ch.
Swell/ Dip/ Int/ Transient/
Inrush
new event is added and
displayed on the LCD
Lo is kept for 1 sec, Hi is
restored
Unbalance rate
exceeding preset threshold
------
Output format : Open collector output
Max input : 30V, 50mA, 200mW
Output voltage : Hi – 4 ~ 5V
Lo – 0 ~ 1V
+5V
H
L
OUTPUT