Setting for vt ratio – KYORITSU 6310 Instruction Manual User Manual
Page 47
Advertising

KEW6310 4.2.1 Basic Setting
KEW6310
4.6
Setting for VT Ratio
0.01 ~ 9999.99 (can be set by 0.01)
* Default value (or after system reset) : 1.00
For the details of VT ratio, refer to “5.4 VT / CT Ratio” in this manual.
1
Press the
Cursor
Keys and select [VT Ratio], and then press the
ENTER
Key.
2
Press the
Cursor
Keys and alter the values, and press the
ENTER
Key
to fix it.
Box with
▲▼
mark appears
at the hundredths' place.
Press the
Cursor
to toggle the
value from 0 to 9.
When increasing a value to 0 in ascending sequence,
value at the tents' place increases by 1.
Press the
Cursor
to toggle the
value from 9 to 0.
Advertising