KEW6310 Safety warnings

KEW6310

8

DANGER


● Verify proper operation on a known source before use.
● Verify proper operation on a known source before use or taking action as a result of the indication of the
instrument.
● Never make measurement on a circuit in which the electrical potential exceeds AC600V.
● Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the
instrument may cause sparking, which can lead to an explosion.
● Never attempt to use the instrument if its surface or your hand are wet.

- Measurement -

● Do not exceed the maximum allowable input of any measuring range.
● Never open the Battery cover and CF card connector cover during a measurement.

- Battery -

● Never open the Battery Cover during a measurement.
● Brand and type of the batteries to be used should be harmonized.

- Power cord -

● Connect the Power cord mains plug to a mains socket outlet
● Use only the Power cord supplied with this instrument.

- Power supply connector -

● Never touch the Power supply connector although it is insulated while the instrument is operating with
batteries.

- Voltage test leads -

● Use only the ones supplied with this instrument.
● Confirm that the measured voltage rating of the test lead is not exceeded.
● Do not connect a Voltage test lead unless required for measuring the parameters desired.
● Connect Voltage test leads to the instrument first, and only then connect them to the circuit under
test.
● Never disconnect Voltage test leads while the instrument is in use.
● Connect to the downstream side of a circuit breaker since a current capacity at the upstream side is large.
● Do not touch two lines under test with the metal tips of the test leads.
● Never touch the metal tips of the test leads.

- Clamp sensor -

● Use only the ones dedicated for this instrument.
● Confirm that the measured voltage rating of the test lead is not exceeded.
● Do not connect a Camp sensor unless required for measuring the parameters desired.
● Connect sensors to the instrument first, and only then connect them to the circuit under test.
● Never disconnect sensors while the instrument is in use.
● Connect to the downstream side of a circuit breaker since a current capacity at the upstream side is large.
● Do not touch two lines under test with the metal tips of the test leads.