KEW6310 16.4 Demand measurement

KEW6310

16.6

16.4 Demand measurement ( Range)

(1) Target value (DEM Target)

Display range

Fixed set value (1.000mW ~ 999.9TW)

(2) Predicted value (DEM Guess)

Display range

Same decimal point place and unit to target value

Formula

×

=

interval

demand

of

beginning

from

Period

interval

Demand

DEM

DEM

GUESS

_

_

_

_

_

_

(3) Demand value (present value) (ΣDEM)

Display range

Same decimal point place and unit to target value

Formula

(

)

interval

hour

WP

DEM

1

×

+

=

Σ

where:

Σ

=

Σ

DEMi

DEM

*

1

=

i

(1P2W×1)

*

2

=

i

(1P2W×2,1P3W,3P3W)

*

3

=

i

(1P2W×3,3P3W3A,3P4W)

*

4

=

i

(1P2W×4)

(4) Load factor

Display range

0.00 ~ 9999.99%

Formula

et

T

DEM

DEM

arg

Σ

16.5 Waveform measurement ( Range)

Displayed data

2 waveforms (256 points)

Scale change

0.1/ 0.2/ 0.5/ 1.0/ 2.0/ 3.0 times of rating

16.6 Harmonic measurement ( Range)

Meas. Method

PLL synchro system

Measuring range

45 ~ 65Hz

Analysis order

1 ~ 63rd

Window width

2 cycles

Window type

Rectangular

Analysis data

512 points

Analyzing rate

approx once / 2 sec

Display item

(1) Voltage, current, THD, frequency
(2) Voltage/ Rate of content/ Phase angle at each order

Save item

(1) Voltage, current, THD
(2) Voltage/ Phase angle at each order