Setting for trigger point – KYORITSU 6310 Instruction Manual User Manual
Page 85
Advertising

4.2.2 Measurement Setting (QUALITY-Swell, Dip, INT Measurement) KEW6310
KEW6310
4.44
Setting for trigger point
Trigger to start and stop recording, when a preset threshold is exceeded, is decided based on the number of
recorded data.
Past: 0 ~ 200 (can be set by 1)
Next : 200 ~ 0 (can be set by 1)
* Default value (or after system reset) : 100
Example of Trigger Pint Setting:
Setting item
e.g.
Reference voltage
100V
Swell 110%
Hysteresis 1%
Trigger point
Past: 100, Next: 100
Start of swell
End of swell
100%
(reference
voltage
110%
(Swell)
109%
(110% - 1%)
total 201 data pts
total 201 data pts
100
data pts
100
data pts
Trigger
Point
100
data pts
100
data pts
Trigger
Point
Duration