Transvenous Pacemaker Testing

Setup

and

Testing

4

4-15

F1

F2

F3

F4

eyr038.eps

6.

Press

F2 RATE

to toggle between

NORM

and

<20.

Note

The

RATE

<20 is used to measure pacer activity under the rate of 20

PPM. The sampling time is extended to accommodate this lower rate.

7. Press

F4 LOAD

UP

or

DOWN

arrow key to select the desired load for

the selected channel.

8. Press

F1

HOLD

to advance the screen:

F1

F2

F3

F4

eyr016.eps

9. Press

F4

XMIT DATA

to transmit data through the serial port.

10. Press

F1

RELEASE

to release the hold and return to the

HOLD

screen.