Fluke Biomedical SigmaPace 1000 User Manual
Page 107
Advertising
![background image](/files/785226/content/doc107.png)
Transvenous Pacemaker Testing
Setup
and
Testing
4
4-21
The screen advances for wave and width selection (the example is the
atrial channel).
F1
F2
F3
F4
eyr070.eps
4. Press
F3
WAVE
UP
or
DOWN
arrow key to select a waveform.
5. Press
F4
WIDTH
UP
or
DOWN
arrow key to adjust the width of the
waveform.
6. Press
F1