SigmaPace 1000

Operators Manual

5-22

RDDCLEAK

This command queries the Analyzer for the currently active dc leakage test
reading and the Analyzer returns the data in the following format:

DCLEAK=+XX.X A.

Error Messages

Error messages are returned instead of data if the required test conditions are
not fulfilled.

During STATIC tests, an error message PULSE is returned if a pacer

pulse output is present on the Analyzer input terminals during this
“power off” continuous test.

During the atrial and the ventricular channel DYNAMIC tests:

An error message PULSE is returned if pulse output is present on the
Analyzer input terminals during these “power on” triggered tests.
This error message is not active during the DCLEAK=DYNAMICAV
test since no discernable pulse should be present across the two A (+)
and V (+) pacemaker channel outputs during the DDD mode of
operation.

An error message NO PULSE is returned if no pacer pulse output is
present on the Analyzer input terminals during these two “power on”
triggered tests. This error message is not active during the
DCLEAK=DYNAMICAV test, since no discernable pulse should be
present across the two A (+) and V (+) pacemaker channel outputs
during the DDD mode of operation.

SETTEST=DCLOADI

The Analyzer measures the dc current present at its associated input terminals.
This dc current can be input in either a positive (+) or negative (-) polarity.

RDDCLOADI

This command queries the Analyzer for the current dc load current test reading
and the Analyzer returns the data in the following format:

DCLOADI=+XX.X mA