Introduction and Specifications

Instrument

Specifications

1

1-27

Battery Test Fixture ............................................ Includes the 9 volt battery supply and

facilitates connection of Analyzer to
the recessed battery terminals within
the Medtronic 5388 and 5348
Temporary Pacemakers.

Long Term Test

Test Configuration................................................ Atrial Channel or Ventricular Channel
Pulse Count Range............................................... 999,999 (maximum)
Rate ........................................................................ 2 % to 20 % (10 % default)
Amplitude .............................................................. 2 % to 20 % (10 % default)
Test Time (max) .................................................... 999:59:59 (hhh:mm:ss)
Maximum error Count .......................................... 200
Test Termination................................................... Manual or when maximum error count

is sensed.

Testloads Selectable ............................................ 200 , 500 , or 1000

Interactive Pacer ECG Simulation

Simulates demand, continuous, non-capture, and non-function patient ECG
activity.

Additional user-selectable parameters

PR Interval .......................................................... 0.05 to 0.300 Seconds
Independent Capture / threshold ........................ 1 mA to 25 mA (1 mA steps) for each

pacemaker channel.

Available Testloads (500

Ω

Default)

Atrial Channel

Selections ........................................................... 200

Ω

, 500

Ω

, or 1000

Ω

Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts

Ventricular Channel

Selections ........................................................... 200

Ω

, 500

Ω

, or 1000

Ω

Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts
Tracking .............................................................. Atrial and ventricular channel settings

are identical.