Fluke Biomedical SigmaPace 1000 User Manual

Page 83

Advertising
background image

Transcutaneous Pacemaker Testing

Setup

and

Testing

3

3-29

3. For each, pulse rate and amplitude, press

F3 LIMIT

UP

or

DOWN

arrow

key to select the limit value.

4. Press

F4

LOAD UP

or

DOWN

arrow key to select the load.

5. Press

F1

NEXT

to advance the screen.

F1

F2

F3

F4

eyr027.eps

6. Press

F3

AMP

UP

or

DOWN

arrow key to adjust the amplitude.

7. Press

F4

RATE

UP

or

DOWN

arrow key to adjust the rate.

Note

Target values for amplitude and rate must be entered for the test to
perform correctly.

8. Press

F1

START

to advance the screen and start the test.