Fluke Biomedical SigmaPace 1000 User Manual

Page 121

Advertising
background image

Transvenous Pacemaker Testing

Setup

and

Testing

4

4-35

c. Remove the battery from the pacemaker and slide the black block

(with the two banana plugs) into the empty battery compartment.

d. Apply light pressure to the pacemaker’s battery compartment door to

seat the test fixture to the recessed battery terminals.

e. To test other brands and/or models of transvenous pacemakers, just

attach test leads (not supplied) between the test fixture banana plugs
and the exposed battery contacts.

3. Press

F1

NEXT

to advance the screen and start the test.

F1

F2

F3

F4

eyr047.eps

4.

Press

F1

HOLD

to advance the screen:

F1

F2

F3

F4

eyr016.eps