Transcutaneous Pacemaker Testing

Setup

and

Testing

3

3-15

F1

F2

F3

F4

eyr017.eps

4. Press

F2

SELECT DRIVE<

to toggle between the Overdrive and

Underdrive BPM rates.

5. Once a rate is selected, press

F4

DRIVE RATE BPM

UP

or

DOWN

arrow key to adjust that rate.

6. Press

ESC

to return to previous menu(s).

Continuous Mode Pacemaker’s Interaction with ECG
Signal

The following test verifies the continuous (or non-demand) pacemaker’s
ability to interact with a simulated ECG signal.

F1

F2

F3

F4

eyr014.eps