Introduction and Specifications

Instrument

Specifications

1

1-25

Resolution ............................................................. ±1 LSD
Display Format...................................................... 3 digits

Physiological Simulation
Selection (Default) .............................................. Single Channel Simulation (V)

Square Wave

Ventricular Channel ............................................ Simulated R-wave
Width................................................................... 1.0 mS
Amplitude............................................................ 20 mV

peak

Active Outputs .................................................... Ventricular Channel

(4 mm Banana Jacks) Only

Available Test Load(s) ........................................ 500

Ω

Only

Additional Waveform Selections

Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width2 mS rise time / 13 mS fall time)

Amplitude

Pacer Load Selection.......................................... 500

Ω

Range ............................................................. 0.05 mV

peak

(50 V

peak

) to

50.0 mV

peak

Accuracy......................................................... ±5 % of setting
Resolution (Step Size).................................... 0.05 mV

peak

steps from 0.05 mV

peak

to

0.95 mV

peak

0.50 mV

peak

steps from 1.0 mV

peak

to

50.0 mV

peak

Default Setting ................................................ 20.0 mV

peak

Width

Range ................................................................. 0.15 mS to 300 mS
Accuracy ............................................................. 5 % of setting
Selection Count .................................................. 58
Resolution (Step Size)........................................ 0.05 mS steps from 0.15 mS to

0.95 mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS

Default Setting (mS) ........................................... 30 mS
Intended Pacemaker Types................................ VVI (Ventricular Pace and Sense

Only)

Compatible Pacemaker Rates ............................ 20 to 200 PPM