Transcutaneous Pacemaker Testing

Test

Options

3

3-3

Test Options

The scheme for transcutaneous pacemaker testing is shown in Figure 3-1. The
specified (or default) test load is pre-selected with each algorithm. If desired,
you can easily select any test load (50

Ω

to 1550

Ω

in 50

Ω

steps) with each

measurement algorithm.

Ensure that you select a test load within the device manufacturer’s specified
operational range. Device-specific measurement algorithms are selectable for
the following list of manufacturers:

Physio-Control Lifepak Series (50

Ω

)

HP / Agilent Codemaster (400

Ω

)

GE / Marquette Responder Series (300

Ω

)

Medical Data Electronics (600

Ω

)

Medical Research Laboratories (500

Ω

)

Zoll Medical PD-Series (1000

Ω

)

Zoll Medical M-Series (100

Ω

)

Tests requiring ECG simulation have three available active outputs:

Low Level (2): 5 Lead: RL / N, RA / R, LA / L, LL / F, and V1 / C1

Transthoracic: Across the selected ventricular test load

High Level (1): Output Amplitude tracks low level Lead I (1000:1)

+ / - 5.0V-peak maximum amplitude
Connector Type: Subminiature Phone Jack