Remote Operation
Transvenous Pacemaker Remote Setup and Testing
5
5-23
The polarity of the dc current is indicated with either positive (+) or negative
(-) sign.
Readings are available in milliamperes as specified, if the input dc current
exceeds the maximum response of: DCLOADI= XXX mA to the host
controller.
EXITTEST
Exits the selected test measurement.
EXITMODE
Halts all tests and returns the Analyzer to main menu serial control.