Error codes – Fluke Biomedical SigmaPace 1000 User Manual
Page 135

Remote Operation
Introduction
5
5-5
REMOTE
The Analyzer enters the remote mode of operation.
All serial commands that do not return data or a predetermined response send
OK to indicate that the serial command was received, understood and
executed.
LOCAL
The Analyzer enters the local (manual) mode of operation.
QBAUD
Diagnostic; the baud rate setting of the Analyzer is transmitted to the host
controller in response to this serial command.
BAUD=nnnn (Only baud rate settings of 2400, 9600, or 19200 are valid).
Error Codes
The following error codes are returned in response to commands that cannot be
executed.
ERR=00
No commands allowed at this time
ERR=01
Unknown command
(Returns the question mark character (?) to the host controller via the serial
port.)
ERR=02
Illegal command. The command was recognized, but not allowed at this time