Transcutaneous Pacemaker Testing

Setup

and

Testing

3

3-31

10. Press

F3 REVIEW UP

or

DOWN

arrow key to view deviations falling

outside the selected deviation values for rate and amplitude.

Up to 200 deviations will be stored in the Analyzer memory. Any long
term test exceeding 200 deviations is automatically terminated. The pulse
width at the “out of limit” pacemaker pulse is also reported.

11. Press

F4 XMIT

to transmit data.

12. Press

F1 RESTART

to restart the test.

13. Press

ESC

to return to previous menu(s).