Transvenous Pacemaker Testing

Setup

and

Testing

4

4-19

F1

F2

F3

F4

eyr040.eps

4. Press

F2

SELECT DRIVE<

to toggle between the Over and Under BPM

rates.

5. Once a rate is selected use

F4

DRIVE RATE BPM

UP

or

DOWN

arrow

key to adjust that rate.

6. Press

F1

HOLD

to advance the screen:

F1

F2

F3

F4

eyr016.eps

7.

Press

F4

XMIT DATA

to transmit data through the serial port.

8.

Press

F1

RELEASE

to release the hold and return to the

HOLD

screen.

9.

Press

ESC

to return to previous menu(s).