Paced refractory period test (prp), Sensed refractory period test (srp) – Fluke Biomedical SigmaPace 1000 User Manual

Page 30

Advertising
background image

SigmaPace 1000

Operators Manual

1-16

Active Outputs

Selected Ventricular Channel Test load
Five lead ECG

Testload Amplitude Output

Ventricular Channel

Range..............................................................0.00 (OFF) to 10.0 mV

peak-to-peak

Accuracy..........................................................±5 % of setting
Resolution (Step Size).....................................0.50 mV

peak-to-peak

steps

Pacer Load Range ..........................................(30) 50

Ω

to 1550

Ω

1 %

Five Lead ECG Output

Range..................................................................0.00 (OFF) to 10.0 mV

peak-to-peak

Accuracy..............................................................±5 % of setting
Resolution (Step Size).........................................0.50 mV steps
Calibration Reference..........................................Lead I (RA to LA)

Paced Refractory Period Test (PRP)

Range......................................................................20 mS to 500 mS

Accuracy..............................................................5 % of reading or 1 mS whichever is

greater

Physiological Simulation

Selection..............................................................Triangle (TRI) Wave
Pulse Width .........................................................40 mS
Rate.....................................................................Single pulse: Interactive with applied

pacemaker pulse activity in the
demand mode of operation.

Amplitude ...............................................................1.0 mV

peak

Lead I

Active Outputs .....................................................Five Lead ECG and selected

ventricular channel test load This
waveform is present on both outputs.

Pacemaker Compatibility

Compatible Pulse Rates......................................30 to 200 BPM
Intended Pacemaker Type ..................................VVI (Ventricular Pace and Sense

Only)

Sensed Refractory Period Test (SRP)

Range......................................................................15 mS to 500 mS

Accuracy..............................................................±5 % of reading or ±1 mS, whichever

is greater

Physiological Simulation

Selection..............................................................Triangle (TRI) Wave