Transvenous pacemaker remote setup and testing – Fluke Biomedical SigmaPace 1000 User Manual

Page 143

Advertising
background image

Remote Operation

Transvenous Pacemaker Remote Setup and Testing

5

5-13

zz= % test limit for rate.

nnn= Set point for rate (1PPM resolution.)

The data is returned in the following format;

AMP=#XXX mA, RATE=#XXX PPM, WIDTH=XXX mS when a reading is
out of the prescribed test limit range. For each parameter, the Analyzer returns
a three-digit numeric value (XXX) with a floating decimal point to cover the
multiple ranges. The unit of measure remains the same for all transmitted
readings (mA, PPM, and mS). The Analyzer marks the out-of-limit readings
with the pound (#) sign.

TRENDEND

Halts the trend test.

EXITTEST

Leaves the selected test measurement.

EXITMODE

Halts all tests and returns the Analyzer to the main serial control menu.

Transvenous Pacemaker Remote Setup and
Testing

The following commands are associated with setting up and testing a
transvenous pacemaker in remote mode.

The synchronized measurement of the pacemaker’s atrial and ventricular
output (amplitude, rate, width, energy activity) can be easily accomplished
using the Analyzer in a dual-channel mode. The remaining pacemaker tests
(demand, async, sensitivity, trend) or the setting of specific adjustments (test
load) must be conducted for only one channel at a time.

SETMODE

SETMODE=TVPACEA (atrial channel only)