Fluke Biomedical SigmaPace 1000 User Manual
Page 68
Advertising
![background image](/files/785226/content/doc068.png)
SigmaPace 1000
Operators Manual
3-14
Demand Mode Pacemaker’s Interaction with ECG
Signal
The following test verifies the demand mode pacemaker’s ability to interact
with a simulated ECG signal.
F1
F2
F3
F4
eyr014.eps
1. From the
SELECT TEST menu
, press
F2
DMAND MODE
.
F1
F2
F3
F4
eyr058.eps
2. Press the
F4
UP
or
DOWN
arrow key to select the desired load.
3. Press
F1 START
to advance the screen and start the test.