Noise immunity test, Refractory period test (atrial channel) – Fluke Biomedical SigmaPace 1000 User Manual

Page 37

Advertising
background image

Introduction and Specifications

Instrument

Specifications

1

1-23

Noise Immunity Test

Test Type ............................................................... Qualitative
Channels

Single.................................................................. Atrial or Ventricular Only

Waveform .............................................................. Sine Wave
Frequency.............................................................. 50 and 60 Hz
Accuracy................................................................ 0.5 Hz
Active Output(s).................................................... Selected atrial and / or ventricular

channel test load

Output Selections

Atrial Channel Only
Ventricular Channel Only

ECG Signal ............................................................ ECG Signal can be added to the

selected Channel

Amplitude

Pacer Load Selection.......................................... 500

Ω

Range ............................................................. 0.00 (OFF) to 100 mV

peak-to-peak

Accuracy......................................................... ±5 % of setting
Resolution....................................................... 5 mV

peak-to-peak

steps

Pacer Load Selection.......................................... 200

Ω

Range ............................................................. 0.00 (OFF) to 40 mV

peak-to-peak

Accuracy......................................................... ±5 % of setting
Resolution....................................................... 5 mV

peak-to-peak

steps

Pacer Load Selection.......................................... 1000

Ω

Range ............................................................. 0.00 (OFF) to 200 mV

peak-to-peak

Accuracy......................................................... ±5 % of setting
Resolution....................................................... 5 mV

peak-to-peak

steps

Refractory Period Test (Atrial Channel)

Test Selections

Paced Refractory Period (PRP)
Sensed Refractory Period (SRP)

Period..................................................................... 20 to 500 mS
Accuracy................................................................ ±5 % of reading or ±1 mS whichever is

greater

Resolution ............................................................. ±1 LSD
Display Format...................................................... 3 digits