SigmaPace 1000

Operators Manual

5-16

RDMETERAVI

This command is compatible with dual-chamber AV sequential pacemakers
only. The Analyzer transmits the following information string for the AV
interval (or delay) from the atrial and the ventricular data “frozen” from the
above HOLD serial command.

The data is returned in the following format:

AV Interval: XXX mS

The Analyzer returns a three-digit numeric value (XXX) with a floating point
decimal to cover the multiple ranges. The unit of measure remains the same
for all transmitted readings (mS)

Additionally, single parameter information can be acquired for either of the
channels:

RDAMPA
RDAMPV

The Analyzer transmits the following information string for amplitude from
the data “frozen” with the above HOLD serial command.

The data is returned in the following format:

A Channel: AMP=XXXmA

V Channel: AMP=XXXmA

For each parameter, the Analyzer returns a three-digit numeric value (XXX)
with a floating point decimal to cover the multiple ranges. The unit of measure
remains the same for all transmitted readings (mA)

RDRATEA
RDRATEV

The Analyzer transmits the following information string for rate from the data
“frozen” with the above HOLD serial command.

The data is returned in the following format:

A Channel: RATE=XXXPPM

V Channel: RATE=XXXPPM