Fluke Biomedical SigmaPace 1000 User Manual

Page 144

Advertising
background image

SigmaPace 1000

Operators Manual

5-14

SETMODE=TVPACEV (ventricular channel only)

SETMODE=TVPACEAV (Both atrial and ventricular channels)

Selects the Analyzer’s transvenous pacer test mode of operation.

Note

All subsequent serial port communications must relate to transvenous
pacer features and follow the prescribed measurement protocols
listed below.

SETLOAD

SETLOAD=nnnn (Both atrial and ventricular channels)

This command is used to select the desired test load. The above “nnnn”
represents the actual test load resistance value (in ohms) and is based upon the
available factory preset selections. The default test load value is 500

Ω

for

both channels.

(Example: SETLOAD=0500 is the command for the selection of a 500 ohm
test load for both the atrial and ventricular channels.)

Valid selections for “nnnn” are as follows: 0200, 0500, and 1000

Ω

RDLOAD

Reads the current test load setting. When this command is sent to the
Analyzer, it returns nnnn; where nnnn equals resistance in ohms from 0200 to
1000

Ω

.

SETTEST

SETTEST=PULSE

The Analyzer monitors the applied pulse rate, amplitude, width, and energy
data for both the atrial and the ventricular pacemaker channels.

HOLD

Freezes the pacer measurement and maintains last reading in memory for both
the atrial and the ventricular pacemaker channels.

Advertising