Transvenous Pacemaker Testing

Setup

and

Testing

4

4-41

F1

F2

F3

F4

eyr051.eps

7. Press

F3 AMP

to adjust the amplitude.

8. Press

F4 RATE

to adjust the rate.

9. Press

F1

START

to advance screen and start the test.

Note

Target values for amplitude and rate must be entered for test to
perform correctly.