Transvenous Pacemaker Testing
Setup
and
Testing
4
4-41
F1
F2
F3
F4
eyr051.eps
7. Press
F3 AMP
to adjust the amplitude.
8. Press
F4 RATE
to adjust the rate.
9. Press
F1
START
to advance screen and start the test.
Note
Target values for amplitude and rate must be entered for test to
perform correctly.