SigmaPace 1000

Operators Manual

1-22

Available Selection(s)......................................(3) 200, 500 and 1000

Ω

±1 %

Default Setting.................................................500

Ω

Amplitude

Pacer Load Selection ..........................................500

Ω

Range..............................................................0.05 mV

peak

(50 V

peak

) to 50.0 mV

peak

Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV

peak

steps from 0.05 mV

peak

to

0.95 mV

peak

0.50 mV

peak

steps from 1.0 mV

peak

to

50.0 mV

peak

Pacer Load Selection ..........................................200

Ω

Range..............................................................0.05 mV

peak

(50 V

peak

) to 20.0 mV

peak

Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV

peak

steps from 0.05 mV

peak

to

0.95 mV

peak

0.50 mV

peak

steps from 1.0 mV

peak

to

20.0 mV

peak

Pacer Load Selection ..........................................1000

Ω

Range..............................................................0.05 mV

peak

(50 V

peak

) to 100 mV

peak-to-

peak

Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV

peak

steps from 0.05 mV

peak

to

0. 95 mV

peak

0.50 mV

peak

steps from 1.0 mV

peak

to

49.5 mV

peak

05.0 mV

peak

steps from 50 mV

peak

to

100 mV

peak

Default Setting .......................................................2.5 mV

peak

Width

Range..................................................................0.15 mS to 300 mS
Accuracy..............................................................±5 % of setting
Selection Count ...................................................58
Resolution (Step Size).........................................0.05 mS steps from 0.15 mS to 0.95

mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS

Intended Pacemaker Type(s)

VVI (Atrial Pace and Sense Only)
Compatible Pulse Rates......................................30 to 200 PPM