Static tests (pacemaker power off) – Fluke Biomedical SigmaPace 1000 User Manual

Page 94

Advertising
background image

SigmaPace 1000

Operators Manual

4-8

ECG

Sensing
& Pacer

Control

Trans

v

eno

u

s

Pacemaker

Static &
Dynamic

DC

Leakage

V

entric

u

lar

Channel

Po

w

er

O

N

/OFF

Atrial Channel

A+

A+

A-

V

+

V

+

V

-

SIGMAPACE 1000
External Pacemaker Analyzer

L
O
A
D
S

L
O
A
D

M

L
O
A
D
S

eyr006.eps

Figure 4-2. DC Leakage Testing

Potentially harmful dc levels can be detected using the various test
configurations described below:

Static Tests (Pacemaker Power OFF):

These tests continuously measure dc leakage as follows:

Atrial Channel: dc leakage across the atrial channel 500

Ω

test load.

Ventricular Channel: dc leakage across the ventricular channel 500

Ω

test load.

AV Channel: dc leakage across an auxiliary 500

Ω

test load

connected to the +atrial and +ventricular outputs.