Fluke Biomedical SigmaPace 1000 User Manual
Page 106
Advertising
![background image](/files/785226/content/doc106.png)
SigmaPace 1000
Operators Manual
4-20
Demand Mode Pacemaker’s Ability to Sense ECG
Activity
The two following, related tests, PRP and SRP, determine the demand mode
pacemaker’s ability to sense ECG activity immediately following either a
paced event (PRP) or a sensed ECG event (SRP).
F1
F2
F3
F4
eyr033.eps
1. From the
SELECT TEST
menu, press
F1
REFRACT DEMAND
.
F1
F2
F3
F4
eyr069.eps
2. Select either of the two options:
•
F1
ATR
for the atrial channel.
•
F2 VENT
for the ventricular channel.