Demand Mode Pacemaker’s Ability to Sense ECG
Activity

The two following, related tests, PRP and SRP, determine the demand mode
pacemaker’s ability to sense ECG activity immediately following either a
paced event (PRP) or a sensed ECG event (SRP).

F1

F2

F3

F4

eyr033.eps

1. From the

SELECT TEST

menu, press

F1

REFRACT DEMAND

.

F1

F2

F3

F4

eyr069.eps

2. Select either of the two options:

F1

ATR

for the atrial channel.

F2 VENT

for the ventricular channel.