Fluke Biomedical SigmaPace 1000 User Manual

Page 145

Advertising
background image

Remote Operation

Transvenous Pacemaker Remote Setup and Testing

5

5-15

RDA
RDV

The Analyzer transmits the following information for rate, amplitude, and
width from the atrial or the ventricular data “frozen” from the above HOLD
serial command. The data is returned in the following format:

AMP=XXX mA

RATE=XXX PPM

WIDTH=XXX mS

For each parameter, the Analyzer returns a three-digit numeric value (XXX)
with a floating point decimal to cover the multiple ranges. The unit of measure
remains the same for all transmitted readings (mA, PPM, and mS).

RDAH

Returns atrial amplitude and rate.

RDVH

Returns ventricular amplitude and rate.

RDAV

The Analyzer transmits the following information string for rate, amplitude,
and width from the atrial or the ventricular data “frozen” from the above
HOLD serial command.

The data is returned in the following format:

A Channel: AMP=XXX mA, RATE=XXX PPM, WIDTH=XXX mS

V Channel: AMP=XXX mA, RATE=XXX PPM, WIDTH=XXX mS

AV Interval: XXX mS

For each parameter, the Analyzer returns a three-digit numeric value (XXX)
with a floating point decimal to cover the multiple ranges. The unit of measure
remains the same for all transmitted readings (mA, PPM and mS)