Long term test, Setup and testing – Fluke Biomedical SigmaPace 1000 User Manual

Page 97

Advertising
background image

Transvenous Pacemaker Testing

Setup

and

Testing

4

4-11

Standard test leads or short jumpers can be used to make connections on other
brands and models of pacemakers.

Long Term Test

Tests are performed for only one channel at a time. The overall stability of the
pacemaker’s pulse rate and amplitude are monitored continuously over an
extended time period. The user can independently input desired test limits for
amplitude and rate (2-20) or use the factory default 10 % limits. Upon
completion of the desired long-term test duration, the user can review
pacemaker output pulses falling outside these test limits.

Setup and Testing

The following describes how to set up and test a transvenous pacemaker.

F1

F2

F3

F4

eyr001.eps

1. From the

Select external pacemaker type

screen, press

F2 INV

to

begin Invasive (transvenous) pacemaker testing.

Advertising