Refractory period test (ventricular channel), Sigmapace 1000 – Fluke Biomedical SigmaPace 1000 User Manual

Page 38

Advertising
background image

SigmaPace 1000

Operators Manual

1-24

Physiological Simulation

Selection (Default)...............................................Single Channel Simulation (A)

Square Wave

Atrial Channel......................................................Simulated P-wave
Width ...................................................................1.0 mS
Amplitude ............................................................20.0 mV

peak

Active Outputs .....................................................Atrial Channel (4 mm Banana Jacks)

Only

Additional Waveform Selections

Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width 2 mS rise time / 13 mS fall time)

Amplitude

Range...................................................................05 mV

peak

(50 V

peak

) to 50.0 mV

peak

Accuracy..............................................................±5 % of setting
Resolution (Step Size).........................................0.05 mV

peak

steps: 0.05 mV

peak

to 0.95

mV

peak

0.50 mV

peak

steps: 1.0 mV

peak

to 49.5

mV

peak

Width

Range..................................................................0.15 mS to 95.0 mS
Accuracy..............................................................±5 % of setting
Selection Count ...................................................50
Resolution (Step Size).........................................0.05 mS steps: 0.15 mS to 0.95 mS

1.0 mS steps: 1.0 mS to 19.0 mS
5.0 mS steps: 20 mS to 95.0 mS

Active Output.......................................................Atrial Channel (4 mm Banana Jacks)

Only

Intended Pacemaker Types ................................AAI (Atrial Pace and Sense Only)
Compatible Pacemaker Rates.............................30 to 200 PPM
Available Test Load.............................................500

Ω

±1 %

Refractory Period Test (Ventricular Channel)

Test Selections

Paced Refractory Period (PRP)
Sensed Refractory Period (SRP)

Period .....................................................................20 to 500 mS
Accuracy ................................................................±5 % of reading or 1 mS whichever is