Asynchronous mode test – Fluke Biomedical SigmaPace 1000 User Manual

Page 28

Advertising
background image

SigmaPace 1000

Operators Manual

1-14

Auxiliary Rate Range

Underdrive (minimum).........................................10 BPM
Overdrive (maximum)..........................................300 BPM

Active Outputs

Five lead ECG
Selected Ventricular Channel Test Load
High Level ECG Jack
This waveform is present on these three outputs.

Pacemaker Compatibility

Compatible Pulse Rates......................................30 to 200 PPM
Intended Type .....................................................External Transcutaneous VVI

(Ventricular Only Pace and Sense)

Asynchronous Mode Test

Test Type................................................................Qualitative
Physiological Simulation Selection

Selection..............................................................Normal Sinus Rhythm (NSR)

generates the complete P-QRS-T
complex.

Amplitude (V

peak

)..................................................1.0 mV Lead I

Rate (Interval)

Underdrive...........................................................NSR @ 85 % of measured pulse

interval / rate

Overdrive.............................................................NSR @ 115 % of measured pulse

interval / rate

Auxiliary Control ..................................................The under / overdrive NSR simulations

can be independently adjusted in 1
BPM increments.

Auxiliary Rate Range

Underdrive (minimum).........................................10 BPM
Overdrive (maximum)..........................................300 BPM

Active Outputs

Five lead ECG
Selected Ventricular Channel Test Load
This waveform is present on both outputs.

Pacemaker Compatibility

Compatible Pulse Rates......................................30 to 200 PPM

Intended Type

External Transcutaneous
VOO (Asynchronous Ventricular Only Pace)