Remote Operation

Transvenous Pacemaker Remote Setup and Testing

5

5-21

uu= % test limit for amplitude.

nnnn= Set point for amplitude (0.1 mA resolution.)

zz= % test limit for rate.

nnn= Set point for rate (1 PPM resolution.)

The data is returned in the following format when a reading is out of the
prescribed test limit range. For each parameter, the Analyzer returns a three-
digit numeric value (XXX) with a floating decimal point to cover the multiple
ranges. The unit of measure remains the same for all transmitted readings
(mA, PPM and mS). The Analyzer marks out-of-limit readings with the pound
(#) sign.

A Channel: AMP=#XXX mA, RATE=#XXX PPM, WIDTH=XXX mS

TRENDEND

Halts the trend test.

EXITTEST

Leaves the selected test.

SETTEST=DCLEAK

The basic dc leakage measurement function is activated when this serial
command is received by the Analyzer. The following complementary serial
commands are then used to select the specific measurement type and the input
configuration of the Analyzer’s dc micro-amp meter function.

DCLEAK=STATICA Continuous measurement on the atrial inputs.

DCLEAK=STATICV Continuous measurement on the vent inputs.

DCLEAK=STATICAV Continuous measurement on the AV (+) inputs.

DCLEAK=DYNAMICA Gated measurement on the atrial inputs.

DCLEAK=DYNAMICV Gated measurement on the vent inputs.

DCLEAK= DYNAMICAV Gated measurement on the AV (+) inputs.