Transcutaneous Pacemaker Testing

Setup

and

Testing

3

3-27

F1

F2

F3

F4

eyr025.eps

2. Press

F1 OUTPUT

to toggle among four simulated operational modes:

Demand

Continuous

Non-capture

Non-function.

3. Press

F3 THRESH

UP

or

DOWN

arrow key to adjust the threshold

amplitude.

4. Press

F4 NSR RATE UP

or

DOWN

arrow key to adjust the NSR rate.

Note

The NSR rate function is active only during the demand mode of
operation.

Long Term Tests

The overall stability of the pacemaker’s pulse rate and amplitude are
monitored continuously over an extended period of time. The user enters the
desired target values for these two parameters into the Analyzer. This program