Transvenous pacemaker tests – Fluke Biomedical SigmaPace 1000 User Manual

Page 31

Advertising
background image

Introduction and Specifications

Instrument

Specifications

1

1-17

Pulse Width......................................................... 40 mS
Rate .................................................................... Double pulse: interactive with applied

pacemaker pulse activity in the
demand mode of operation.

Amplitude .............................................................. 1.0 m V

peak

Lead I

Active Outputs ...................................................... Five Lead ECG and selected

ventricular channel test load. This
waveform is present on both outputs.

Pacemaker Compatibility

Compatible Pulse Rates ..................................... 30 to 200 BPM
Intended Pacemaker Type.................................. VVI (Ventricular Pace and Sense

Only)

Transvenous Pacemaker Tests

Output Pulse Measurement

Test Type ............................................................... Quantitative
Display Formats

Single.................................................................. Atrial or Ventricular only
Dual .................................................................... Atrial and ventricular Channels

Measurement Ranges

Current................................................................ 0.05 mA to .999 mA* (Single Channel

Only)
1.00 mA to 9.99 mA
10.0 mA to 30.0 mA

Accuracy......................................................... 2 % of reading or ±50 µA (whichever is

greater)

Polarity Indicator ................................................. + or -
Rate .................................................................... 10 PPM to 99.9 PPM

100 PPM to 999 PPM

Accuracy......................................................... 0.5 % or 0.3 PPM (whichever is

greater)

Width................................................................... 0.02 mS to .999 mS

1.00 mS to 9.99 mS
10.0 mS to 99.99 mS

Accuracy......................................................... 0.5 % or ±14 µS (whichever is greater)

Resolution........................................................... ±1 LSD or ±4 µS (whichever is

greater)