SigmaPace 1000

Operators Manual

4-18

Continuous Mode and Dual-channel (AV) ECG Signal

The following test measures the continuous (or non-demand) pacemaker’s
ability to interact with a simulated dual-channel (AV) ECG signal.

F1

F2

F3

F4

eyr032.eps

1. From the

SELECT TEST

menu, press

F3

ASYNC MODE

.

F1

F2

F3

F4

eyr068.eps

2. Press

F4 LOAD

UP

or

DOWN

arrow key to select the load.

3. Press

F1

START

to advance the screen and start the test.